Working principle of ethylene oxide sterilizer
Ethylene oxide sterilizer is a special equipment that uses ethylene oxide gas to fumigate and sterilize items enclosed in the sterilization chamber at a certain temperature, pressure, and humidity. Ethylene oxide sterilization is
Through its non-specific alkylation with microorganisms, the reaction base of microorganisms is suppressed
The normal function of the mouse causes its metabolism to become impaired and die.
2.
Features of ethylene oxide sterilizer
Ethylene oxide gas sterilization can be carried out under lower temperature, pressure and humidity conditions,
It has the characteristics of strong penetrating power, broad-spectrum sterilization, thorough sterilization, no corrosion and no damage to articles.
3.
Main technical indicators of ethylene oxide sterilizer
Temperature control range of
ethylene oxide sterilizer: 20~60℃, maximum control error: empty box<±3℃, full load<10℃
Pressure control range of
ethylene oxide sterilizer: -85Kpa~+80Kpa, accuracy level 1, resolution 0.1Kpa
Humidity control range of ethylene oxide sterilizer: 30%RH~85%RH, error <20%RH, resolution 0.1%RH
The humidification frequency of ethylene oxide sterilizer: any.
Control range of sterilant concentration of ethylene oxide sterilizer: 250㎎/L~1200㎎/L
Time control range of ethylene oxide sterilizer: 30min~12h
Maximum power of ethylene oxide sterilizer: <65KW, heating method: electric heating
Leakage rate of ethylene oxide sterilizer: ≤0.1Kpa/min
Ethylene oxide sterilizer power supply: three-phase four-wire 380V±10%, 50Hz
Working environment of
ethylene oxide sterilizer: temperature range: 5~40℃, humidity range: ≤80%RH
Henan Angel Medical Instruments Co., Ltd supply professional eto sterilizer machine, any need please email to info01@cnoilmachine.com